Reaction time-monitored patient-maintained propofol sedation: a pilot study in oral surgery patients

O'Brien, C., Urquhart, C.S., Allam, S., Anderson, K.J., Leitch, J. A., Macpherson, A. and Kenny, G.N.C. (2013) Reaction time-monitored patient-maintained propofol sedation: a pilot study in oral surgery patients. Anaesthesia, 68(7), pp. 760-764. (doi: 10.1111/anae.12291)

Full text not currently available from Enlighten.

Publisher's URL: http://dx.doi.org/10.1111/anae.12291

Abstract

Previous volunteer studies of an effect-site controlled patient-maintained sedation system using propofol have demonstrated a risk of oversedation. We have incorporated a reaction time monitor into the handset to add an individualised patient-feedback mechanism. This pilot study assessed if the reaction time-feedback modification would prove safe and effective in 20 healthy patients receiving sedation while undergoing oral surgery. All patients successfully sedated themselves without reaching any unsafe endpoints. All 20 maintained verbal contact throughout. The mean (SD) lowest peripheral blood oxygen saturation was 98.0 (2.1)% breathing room air. No patient required supplementary oxygen. The mean (SD) maximum effect-site propofol concentration reached was 1.6 (0.5) μg.ml−1. The present system was found to be safe and effective, allowing oral surgery treatment under conscious sedation, but preventing oversedation.

Item Type:Articles (Newspaper Article)
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:O'Brien, Mr Conor and Kenny, Prof Gavin and Urquhart, Craig
Authors: O'Brien, C., Urquhart, C.S., Allam, S., Anderson, K.J., Leitch, J. A., Macpherson, A., and Kenny, G.N.C.
College/School:College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing
Journal Name:Anaesthesia
ISSN:0003-2409
ISSN (Online):1365-2044

University Staff: Request a correction | Enlighten Editors: Update this record