Winata, S. , Hoinkiss, D. C., Keith, G. A. , Al-Wasity, S. M. and Porter, D. A. (2023) Multislice-to-Volume Prospective Motion Correction for Functional MRI Protocols at 7T. ISMRM & ISMRT Annual Meeting & Exhibition 2023, Toronto, ON, Canada, 03-08 Jun 2023.
Text
319582.pdf Restricted to Repository staff only 2MB |
Abstract
7 Tesla MRI provides higher signal-to-noise ratio (SNR) and spatial resolution, but is also more sensitive to motion-induced artefacts. The restricted environment within 7T scanners make non-hardware options for motion correction attractive. In this abstract, we investigate the use of the markerless, image-based, real-time Multislice Prospective Acquisition Correction (MS-PACE) technique for functional MRI protocols at 7T. This multislice scheme allows for a sub-TR motion detection and correction. It is demonstrated that the technique is able to correct longer-term motion components occurring during the acquisition.
Item Type: | Conference or Workshop Item |
---|---|
Status: | Published |
Refereed: | Yes |
Glasgow Author(s) Enlighten ID: | Winata, Mr Steven and Al-Wasity, Mr Salim and Keith, Dr Graeme and Porter, Professor David |
Authors: | Winata, S., Hoinkiss, D. C., Keith, G. A., Al-Wasity, S. M., and Porter, D. A. |
College/School: | College of Medical Veterinary and Life Sciences > School of Infection & Immunity College of Medical Veterinary and Life Sciences > School of Psychology & Neuroscience |
Related URLs: |
University Staff: Request a correction | Enlighten Editors: Update this record