Long term outcome and elasticity of a polyester mesh used for laparoscopic ventral hernia repair

O'Dwyer, P.J., Chew, C. and Qandeel, H. (2022) Long term outcome and elasticity of a polyester mesh used for laparoscopic ventral hernia repair. Hernia, 26(2), pp. 489-493. (doi: 10.1007/s10029-021-02486-6) (PMID:34426878) (PMCID:PMC8382097)

[img] Text
249341.pdf - Published Version
Available under License Creative Commons Attribution.

532kB

Abstract

Background: Repair of a ventral hernia is increasingly being performed by a laparoscopic approach despite lack of good long term follow up data on outcomes. The aim of this study was to examine the long term performance of a polyester mesh and to assess its elastic properties in patients undergoing laparoscopic ventral hernia repair. Methods: All patients being assessed for a ventral hernia repair between August 2011 and November 2013 were placed on a prospective database. Those undergoing laparoscopic repair with a polyester mesh were seen at clinic at one month and one year, while their electronic records were assessed at 34 months (range 24–48 months) and 104 months (range 92–116 months). In addition, CT scans of the abdomen and pelvis performed for any reason on these patients during the follow up period were reviewed by a consultant gastrointestinal radiologist. Mechanical failure testing of the mesh was also performed. Results: Thirty-two of the 100 patients assessed for ventral hernia repair had a laparoscopic repair with a polyester mesh. Nineteen (59%) had CT scans performed during the follow-up period. No recurrence was recorded at 34 months, while three (9.4%) had a recurrence at 104 months. Two had central breakdown of the mesh at 81 and 90 months, while 1 presented acutely at 116 months after operation. Mesh had stretched across the defect by an average of 21% (range 5.7–40%) in nine patients. Mechanical testing showed that this mesh lost its elasticity at low forces ranging between 1.8 and 3.2 N/cm. Conclusion: This study shows that late recurrence is a problem following laparoscopic ventral hernia repair with polyester mesh. The mesh loses it elasticity at a low force. This combined with degradation of mesh seems the most likely cause of failure. This is unlikely to be a unique problem of polyester mesh and further long-term studies are required to better assess this operative approach to ventral hernia repair.

Item Type:Articles
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:O'Dwyer, Professor Patrick
Authors: O'Dwyer, P.J., Chew, C., and Qandeel, H.
College/School:College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing
Journal Name:Hernia
Publisher:Springer
ISSN:1265-4906
ISSN (Online):1248-9204
Published Online:23 August 2021
Copyright Holders:Copyright © 2021 The Authors
First Published:First published in Hernia 26(2): 489-493
Publisher Policy:Reproduced under a Creative Commons License

University Staff: Request a correction | Enlighten Editors: Update this record