Validation and Safety Approval of a Dual-Mode Head Coil for pTx Applications In Vivo at 7 Tesla

Williams, S. , Allwood-Spiers, S., McElhinney, P. , Tao, Y., Foster, J. E., Porter, D. A. and Gunamony, S. (2020) Validation and Safety Approval of a Dual-Mode Head Coil for pTx Applications In Vivo at 7 Tesla. ISMRM & SMRT Virtual Conference & Exhibition, 8-14 August 2020.

[img] Text
223872.pdf - Published Version

417kB

Publisher's URL: https://www.ismrm.org/20/program_files/DP13-01.htm

Abstract

Following regulatory approval of single-transmit MRI at 7 tesla, there is a rapidly growing interest in clinical MRI at this field strength. However, the wider use of diagnostic MRI at 7T will require imaging in parallel-transmit (pTx) mode to reduce B1+ inhomogeneity. Previous work introduced a dual-mode head coil that operates in both transmit modes and this study investigates the use of this coil for the pTx case. It also describes the safety procedure that was followed to ensure safe operation for human scanning using the real-time SAR supervision on a commercial scanner.

Item Type:Conference or Workshop Item
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Gunamony, Dr Shajan and Williams, Dr Sydney and McElhinney, Dr Paul and Foster, Dr John and Porter, Professor David and Allwood-Spiers, Sarah and Tao, Yuehui
Authors: Williams, S., Allwood-Spiers, S., McElhinney, P., Tao, Y., Foster, J. E., Porter, D. A., and Gunamony, S.
Subjects:Q Science > Q Science (General)
R Medicine > R Medicine (General)
T Technology > T Technology (General)
College/School:College of Medical Veterinary and Life Sciences > School of Psychology & Neuroscience
College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing
Research Group:Imaging Centre of Excellence
Published Online:01 August 2020
Copyright Holders:Copyright © 2020 ISMRM
Publisher Policy:Reproduced with the permission of the publisher
Related URLs:

University Staff: Request a correction | Enlighten Editors: Update this record