Williams, S. , Allwood-Spiers, S., McElhinney, P. , Tao, Y., Foster, J. E., Porter, D. A. and Gunamony, S. (2020) Validation and Safety Approval of a Dual-Mode Head Coil for pTx Applications In Vivo at 7 Tesla. ISMRM & SMRT Virtual Conference & Exhibition, 8-14 August 2020.
![]() |
Text
223872.pdf - Published Version 417kB |
Publisher's URL: https://www.ismrm.org/20/program_files/DP13-01.htm
Abstract
Following regulatory approval of single-transmit MRI at 7 tesla, there is a rapidly growing interest in clinical MRI at this field strength. However, the wider use of diagnostic MRI at 7T will require imaging in parallel-transmit (pTx) mode to reduce B1+ inhomogeneity. Previous work introduced a dual-mode head coil that operates in both transmit modes and this study investigates the use of this coil for the pTx case. It also describes the safety procedure that was followed to ensure safe operation for human scanning using the real-time SAR supervision on a commercial scanner.
Item Type: | Conference or Workshop Item |
---|---|
Status: | Published |
Refereed: | Yes |
Glasgow Author(s) Enlighten ID: | Gunamony, Dr Shajan and Williams, Dr Sydney and McElhinney, Dr Paul and Foster, Dr John and Porter, Professor David and Allwood-Spiers, Sarah and Tao, Yuehui |
Authors: | Williams, S., Allwood-Spiers, S., McElhinney, P., Tao, Y., Foster, J. E., Porter, D. A., and Gunamony, S. |
Subjects: | Q Science > Q Science (General) R Medicine > R Medicine (General) T Technology > T Technology (General) |
College/School: | College of Medical Veterinary and Life Sciences > School of Psychology & Neuroscience College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing |
Research Group: | Imaging Centre of Excellence |
Published Online: | 01 August 2020 |
Copyright Holders: | Copyright © 2020 ISMRM |
Publisher Policy: | Reproduced with the permission of the publisher |
Related URLs: |
University Staff: Request a correction | Enlighten Editors: Update this record