ISCEV extended protocol for VEP methods of estimation of visual acuity

Hamilton, R. , Bach, M., Heinrich, S. P., Hoffmann, M. B., Odom, J. V., McCulloch, D. L. and Thompson, D. A. (2021) ISCEV extended protocol for VEP methods of estimation of visual acuity. Documenta Ophthalmologica, 142(1), pp. 17-24. (doi: 10.1007/s10633-020-09780-1) (PMID:32676804) (PMCID:PMC7906925)

[img] Text
218986.pdf - Published Version
Available under License Creative Commons Attribution.

488kB

Abstract

The International Society for Clinical Electrophysiology of Vision (ISCEV) standard for visual evoked potentials (VEPs) describes a minimum procedure for clinical VEP testing and encourages more extensive testing. This ISCEV extended protocol is an extension to the VEP standard. It describes procedures for recording multiple VEPs to a range of sizes of pattern stimuli to establish the VEP spatial frequency limit (threshold) and for relating this limit to visual acuity.

Item Type:Articles
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Hamilton, Dr Ruth
Authors: Hamilton, R., Bach, M., Heinrich, S. P., Hoffmann, M. B., Odom, J. V., McCulloch, D. L., and Thompson, D. A.
College/School:College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing
Journal Name:Documenta Ophthalmologica
Publisher:Springer
ISSN:0012-4486
ISSN (Online):1573-2622
Published Online:16 July 2020
Copyright Holders:Copyright © 2020 The Authors
First Published:First published in Documenta Ophthalmologica 142(1): 17-24
Publisher Policy:Reproduced under a Creative Commons License

University Staff: Request a correction | Enlighten Editors: Update this record