The accuracy of femoral component rotational measurements using computed tomography - a cadaveric study

Eseonu, O., Cree, C., Sambrook, M., Blyth, M. and Jones, B. (2020) The accuracy of femoral component rotational measurements using computed tomography - a cadaveric study. Anthroplasty, 2, 36. (doi: 10.1186/s42836-020-00052-6)

[img] Text
208206.pdf - Published Version
Available under License Creative Commons Attribution.

1MB

Abstract

Background: CT scans can be used to assess the rotational alignment of the femoral component following total knee arthroplasty (TKA). This is done by calculating the posterior condylar angle (PCA). However, the methods used may not account for the biomechanical functionality of the TKA components. This cadaveric study aimed to determine whether the axis of scanning (mechanical or anatomical) alters the results of PCA calculations. Methods: CT scans of 12 cadaveric adult femora were performed along the anatomical axis and the mechanical axis. The PCA was determined on each CT scan by measuring the relationship of the prosthetic posterior condyles to the surgical epicondylar axis of the femur. The mechanical and anatomical axis groups were further subdivided into best-fit and multi-slice subgroups. As a control, the posterior condylar angle was also calculated on photographic images of each femur. Bland-Altman plots were used to determine the correlation between the PCA values obtained from the different scanning axes and measurement techniques. Results: There was no significant difference between the PCA measurements derived from anatomical and mechanical axis CT scans. The Pearson correlation co-efficient also indicated good correlation between the two scanning axes. Conclusion: The axis of scanning does not significantly affect the PCA measurements. Therefore, the measurements may be reliably used for clinical decision-making, regardless of the axis of CT scanning.

Item Type:Articles
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Eseonu, Mr Onyedikachi and Sambrook, Dr Martin and Jones, Mr Bryn and Blyth, Mr Mark
Authors: Eseonu, O., Cree, C., Sambrook, M., Blyth, M., and Jones, B.
College/School:College of Medical Veterinary and Life Sciences > School of Infection & Immunity
College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing
Journal Name:Anthroplasty
Publisher:BMC
ISSN:2524-7948
ISSN (Online):2524-7948
Copyright Holders:Copyright © 2020 The Authors
First Published:First published in Anthroplasty 2:36
Publisher Policy:Reproduced under a Creative Commons License

University Staff: Request a correction | Enlighten Editors: Update this record