Evaluation of coefficients to derive organ and effective doses from cone-beam CT (CBCT) scans: a Monte Carlo study

Abuhaimed, A. and Martin, C. J. (2017) Evaluation of coefficients to derive organ and effective doses from cone-beam CT (CBCT) scans: a Monte Carlo study. Journal of Radiological Protection, 38, pp. 189-206. (doi: 10.1088/1361-6498/aa9b9f) (PMID:29154259)

[img]
Preview
Text
152785.pdf - Accepted Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

1MB

Abstract

Regular imaging is used throughout image guided radiation therapy (IGRT) to improve treatment delivery. In order for treatment procedures to be optimized, the doses delivered by imaging exposures should be taken into account. CT dosimetry methods based on the CT dose index (CTDI), measured with a 100 mm long pencil ionization chamber (CTDI<sub>100</sub>) in standard phantoms, are not designed for the cone beam CT (CBCT) imaging systems used in radiotherapy, therefore a modified version has been proposed for CBCT by the International Electrotechnical Commission (CTDI<sub>IEC</sub>). Monte Carlo simulations based on a Varian On Board Imaging (OBI) system have been used to derive conversion coefficients that enable organ doses for ICRP reference phantoms to be determined from the CTDI<sub>IEC</sub> for different scan protocols and different beam widths (80 - 320) mm. A dose-width product calculated by multiplying the CTDI<sub>IEC</sub> by the width of the CBCT beam is proposed as a quantity that can be used for estimating effective dose. The variation in coefficients with CBCT beam width has been studied. Coefficients to allow estimation of effective doses have been derived, namely 0.0034 mSv(mGy cm)<sup>-1</sup> for the head, 0.0252 mSv(mGy cm)<sup>-1</sup> for the thorax, 0.0216 mSv(mGy cm)<sup>-1</sup> for the abdomen and 0.0150 mSv(mGy cm)<sup>-1</sup> for the pelvis, and these may be applicable more generally to other CBCT systems in radiotherapy. If data on effective doses are available, these can be used in making judgements on the contributions to patient dose from imaging, and thereby assist in optimization of the treatment regimes. The coefficients can also be employed in converting dosimetry data recorded in patient records into quantities relating directly to patient doses.

Item Type:Articles
Keywords:CT dose index (CTDI), cone beam CT (CBCT), conversion coefficients, image guided radiation therapy (IGRT), varian OBI.
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Martin, Dr Colin
Authors: Abuhaimed, A., and Martin, C. J.
College/School:College of Medical Veterinary and Life Sciences > School of Medicine, Dentistry & Nursing
Journal Name:Journal of Radiological Protection
Publisher:IOP Publishing
ISSN:0952-4746
ISSN (Online):1361-6498
Copyright Holders:Copyright © 2017 IOP Publishing Ltd
First Published:First published in Journal of Radiological Protection 2017
Publisher Policy:Reproduced in accordance with the copyright policy of the publisher

University Staff: Request a correction | Enlighten Editors: Update this record