Gamage, K.A.A. , Joyce, M.J. and Cave, F.D. (2014) Sector-shaped fast organic liquid scintillation detectors based neutron coincidence counter. Applied Radiation and Isotopes, 92, pp. 1-5. (doi: 10.1016/j.apradiso.2014.06.002)
Full text not currently available from Enlighten.
Abstract
A simulation based on the Monte Carlo method is described which has been performed using MCNPX 2.6.0, to model the geometry of a sector-shaped liquid scintillation detector in response to coincident neutron events. The detection of neutrons from a mixed-oxide (MOX) fuel pellet has been simulated for different thicknesses of scintillators. A layer of lead has been used to reduce the gamma-ray fluence reaching the scintillator and, the effect of lead for neutron detection has also been estimated by considering different thicknesses of lead layers.
Item Type: | Articles |
---|---|
Status: | Published |
Refereed: | Yes |
Glasgow Author(s) Enlighten ID: | Gamage, Professor Kelum |
Authors: | Gamage, K.A.A., Joyce, M.J., and Cave, F.D. |
College/School: | College of Science and Engineering > School of Engineering |
Journal Name: | Applied Radiation and Isotopes |
Publisher: | Elsevier |
ISSN: | 0969-8043 |
ISSN (Online): | 1872-9800 |
Published Online: | 12 June 2014 |
University Staff: Request a correction | Enlighten Editors: Update this record