In vitrostudies on the influence of surface modification of Ni-Ti alloy on human bone cells

Chrzanowski, W., Neel, E.A.A., Armitage, D.A., Zhao, X., Knowles, J.C. and Salih, V. (2009) In vitrostudies on the influence of surface modification of Ni-Ti alloy on human bone cells. Journal of Biomedical Materials Research Part A, 93(4), pp. 1596-1608. (doi: 10.1002/jbm.a.32646)

Full text not currently available from Enlighten.

Publisher's URL: http://dx.doi.org/10.1002/jbm.a.32646

Abstract

The in vitro cell behavior on Nitinol™ after different surface treatments was investigated. As references samples, commercially pure titanium (cpTi) and bioactive titanium were used. The surface treatments influenced the topography, surface energy, crystallographic structure, ion release, chemistry, and ability to form apatite layer from simulated body fluids. Regardless of the surface treatment, the bioactivity study showed that the kinetics of apatite film formation was similar for all tested samples. No clear indication of the surface characteristics influence on the ability for calcium-phosphate precipitation was evident. Cell activity studies showed that ground nickel titanium, spark oxidized and thermally oxidized (at 400°C and below) had higher cellular activity and caused increased alkaline phosphatase (ALP) and osteocalcin (OC) expression which was comparable to control tissue culture plastic and titanium reference samples. Regardless of surface modifications, preimmersion of the samples in media for 72 h resulted in cell proliferation at the same level for all samples. Therefore, it can be concluded that preconditioning of samples alters surface properties and modulates the cell response regardless of the initial surface treatment and its properties. Moreover, a detrimental effect on cell response was observed after 7 and 14 days in culture for alkali treated samples. This was attributed to a high surface nickel concentration and a high nickel ion release rate from these surfaces.

Item Type:Articles
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Chrzanowski, Dr Wojciech
Authors: Chrzanowski, W., Neel, E.A.A., Armitage, D.A., Zhao, X., Knowles, J.C., and Salih, V.
College/School:College of Science and Engineering > School of Engineering > Systems Power and Energy
Journal Name:Journal of Biomedical Materials Research Part A
ISSN:1549-3296
ISSN (Online):1552-4965