Scan time reduction for readout-segmented EPI using simultaneous multislice acceleration: Diffusion-weighted imaging at 3 and 7 Tesla

Frost, R., Jezzard, P., Douaud, G., Clare, S., Porter, D. A. and Miller, K. L. (2015) Scan time reduction for readout-segmented EPI using simultaneous multislice acceleration: Diffusion-weighted imaging at 3 and 7 Tesla. Magnetic Resonance in Medicine, 74(1), pp. 136-149. (doi: 10.1002/mrm.25391) (PMID:25078777)

[img]
Preview
Text
198601.pdf - Published Version
Available under License Creative Commons Attribution.

1MB

Abstract

Purpose: Readout‐segmented echo‐planar imaging (rs‐EPI) can provide high quality diffusion data because it is less prone to distortion and blurring artifacts than single‐shot echo‐planar imaging (ss‐EPI), particularly at higher resolution and higher field. Readout segmentation allows shorter echo‐spacing and echo train duration, resulting in reduced image distortion and blurring, respectively, in the phase‐encoding direction. However, these benefits come at the expense of longer scan times because the segments are acquired in multiple repetitions times (TRs). This study shortened rs‐EPI scan times by reducing the TR duration with simultaneous multislice acceleration. Methods: The blipped‐CAIPI method for slice acceleration with reduced g‐factor SNR loss was incorporated into the diffusion‐weighted rs‐EPI sequence. The rs‐ and ss‐EPI sequences were compared at a range of resolutions at both 3 and 7 Tesla in terms of image fidelity and diffusion postprocessing results. Results: Slice‐accelerated clinically useful trace‐weighted images and tractography results are presented. Tractography analysis showed that the reduced artifacts in rs‐EPI allowed better discrimination of tracts than ss‐EPI. Conclusion: Slice acceleration reduces rs‐EPI scan times providing a practical alternative to diffusion‐weighted ss‐EPI with reduced distortion and high resolution. Magn Reson Med 74:136–149, 2015.

Item Type:Articles
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Porter, Professor David
Authors: Frost, R., Jezzard, P., Douaud, G., Clare, S., Porter, D. A., and Miller, K. L.
College/School:College of Medical Veterinary and Life Sciences > School of Psychology & Neuroscience
Journal Name:Magnetic Resonance in Medicine
Publisher:Elsevier
ISSN:0740-3194
ISSN (Online):1522-2594
Published Online:30 July 2014
Copyright Holders:Copyright © 2014 The Authors
First Published:First published in Magnetic Resonance in Medicine 74(1):136-149
Publisher Policy:Reproduced under a Creative Commons License

University Staff: Request a correction | Enlighten Editors: Update this record