Comparison of different diastolic resting indexes to iFR: are they all equal?

Van't Veer, M. et al. (2017) Comparison of different diastolic resting indexes to iFR: are they all equal? Journal of the American College of Cardiology, 70(25), pp. 3088-3096. (doi: 10.1016/j.jacc.2017.10.066) (PMID:29268922)

[img]
Preview
Text
154590.pdf - Published Version
Available under License Creative Commons Attribution Non-commercial No Derivatives.

1MB

Abstract

Background: Pressure measurement for the duration of the wave-free period (WFP) is considered essential for resting-state physiological assessment of coronary stenosis severity using the instantaneous wave-free ratio (iFR). Objectives: The aim of this study was to compare other diastolic resting indexes to iFR. Methods: In the population of the VERIFY2 (Pd/Pa vs iFR in an Unselected Population Referred for Invasive Angiography) study, iFR calculated by proprietary software (Volcano Harvest, Volcano Corporation, Rancho Cordova, California) was compared with the ratio of resting distal coronary pressure and aortic pressure during the complete duration of diastole (dPR), 25% to 75% of diastole (dPR25–75), and midpoint of diastole (dPRmid), along with Matlab calculated iFR (iFRmatlab) and iFR-like indexes shortening the length of the WFP by 50 and 100 ms (iFR−50ms and iFR−100ms), respectively. Mutual differences, Spearman correlations, area under the curve values from receiver-operating characteristic analyses, and diagnostic performance with respect to iFR and fractional flow reserve (FFR) were calculated for all indexes. Results: Median iFR in 197 patients with 257 vessels was 0.91 with an interquartile range of 0.87 to 0.95. The mutual differences (± SD) with iFR were 0.006 ± 0.011 (dPR), 0.001 ± 0.007 (dPR25–75), 0.001 ± 0.008 (dPRmid), 0.005 ± 0.009 (iFRmatlab), 0.003 ± 0.008 (iFR−50ms), and 0.001 ± 0.009 (iFR−100ms). Correlations for all indexes with iFR were >0.99 (p < 0.001 for all). Area under the curve values for predicting iFR were >0.99 for all indexes as well. Diagnostic accuracy compared with FFR was 76% to 77% for all indexes including iFR. Conclusions: All diastolic resting indexes tested were identical to iFR, both numerically and with respect to their agreement with FFR. A numerically equal value to iFR can be determined without restriction to the WFP. Cutoff values, guidelines, and clinical recommendations for iFR can therefore be extended to these other indexes. (Pd/Pa vs iFR in an Unselected Population Referred for Invasive Angiography [VERIFY2]; NCT02377310).

Item Type:Articles
Keywords:FFR, coronary physiology, coronary pressure measurements, iFR, resting indexes, wave-free period.
Status:Published
Refereed:Yes
Glasgow Author(s) Enlighten ID:Hennigan, Dr Barry and Oldroyd, Dr Keith and Berry, Professor Colin
Authors: Van't Veer, M., Pijls, N. H.J., Hennigan, B., Watkins, S., Ali, Z. A., De Bruyne, B., Zimmermann, F. M., van Nunen, L. X., Barbato, E., Berry, C., and Oldroyd, K. G.
College/School:College of Medical Veterinary and Life Sciences > School of Cardiovascular & Metabolic Health
Journal Name:Journal of the American College of Cardiology
Publisher:Elsevier
ISSN:0735-1097
ISSN (Online):1558-3597
Published Online:18 December 2017
Copyright Holders:Copyright © 2017 The Authors
First Published:First published in Journal of the American College of Cardiology 70(25): 3088-3096
Publisher Policy:Reproduced under a Creative Commons License

University Staff: Request a correction | Enlighten Editors: Update this record